Keyence Visual KV Series Manual de usuario Pagina 285

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 444
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 284
7.5 KV Mode Programming
KV-300 Series Only
KV-300
KV-10/80
Chapter 7 KV-L2 Serial Interface Module
1-261
1
7
E2 Program
unregis-
tered
"M1" was transmitted with no
program stored in the KV-300
CPU.
Reading of timer/counter was
attempted.
Register a program into the
KV-300 CPU.
Check the program for the
timer/counter number being
used and specify it.
E3 CPU fault Hardware error was detected
in a KV-300 CPU.
Turn the power to the KV-300
OFF and then ON again. If
the cause of the error is not
corrected, the CPU may be
faulty. Contact your nearest
KEYENCE office.
E4 Write-
protected
Attempt was made to change
the preset value of a counter,
timer, or CTC in a write-
protected program.
Write-enable the program,
and change the preset value.
E5 CPU error "M1" was transmitted before
an error message was
cleared from the KV-300.
Clear the error message from
the KV-300 CPU, and remove
the cause of the error.
Error code list
Error codes included in responses to the KV-300 CPU error check command (?E)
as well as error descriptions, causes of error, and remedies are listed below.
Error
Code Description Cause Remedy
00 No error
10 CALL nest
error
Subroutines are nested to four
levels or more.
These errors may occur when
a program is executed. Check
the program while paying
particular attention to usage of
subroutines and the INT
instruction, and restore the
program accordingly.
11 FOR nest
erro
FOR-NEXT instructions are
nested to eight levels or more.
12 INT nest
error
Five levels or more of inter-
rupts occurred simultaneously.
13 MPS nest
error
MPS/MPP instructions are
nested to eight levels or more.
20 Memory
erased
Contents of RAM are erased.
All data in data memory and
current values of counters in
the KV-300 CPU are cleared.
Check data in data memory
and current values of counters,
and re-enter correct data. The
KV-300 CPU returns to normal
operation after the error is
cleared.
30
Scan time
error
Scan time exceeds 300 ms. Check the program again.
Be sure that there is no
excessive repetition using the
FOR-NEXT loop.
Be sure that an interrupt is not
executed at a higher rate.
40 NO PRO-
GRAM
An attempt was made to
operate the KV-300 CPU with
no program stored.
Write a program into the
KV-300 CPU.
50 SND/RTN
failure
Hardware error was detected
in a KV-300 CPU.
Turn the power to the KV-300
OFF and then ON again. If
the cause of the error is not
removed, the KV-300 CPU
may be faulty. Contact
nearest KEYENCE office.
51 Division by
zero
52 Address
error
53 Invalid
instruction
54 Watchdog
timer error
KVHKA Chap 07_5to10.p65 08.3.11, 11:18 AM261
Vista de pagina 284
1 2 ... 280 281 282 283 284 285 286 287 288 289 290 ... 443 444

Comentarios a estos manuales