LK-G-M-NO3-E
3 Function Settings
Setting the Mutual Interference Prevention Function
When two heads are connected and a little distance is left between them, a measurement error
may occur by receiving the light emitted from the other head. The mutual interference
prevention function is used in such a case.
When the mutual interference prevention function is set to ON, the two heads emit the laser
beam alternately to evade the interference from the other.
Setting procedure of the mutual interference prevention function
1
Hold down the key, and select "
oPt on
" with the key.
Set the common settings.
2
Press the key, and select function
"
b
" with the
key.
Set the mutual interference prevention function.
3
Press the key, and select
"
b-1
"
with the key.
The mutual interference prevention function is set to
"
ON
"
in
this example.
4
Press the key to register the setting. Press the
key to return to the measurement state.
The sampling rate is doubled when the mutual interference prevention function is set to
ON.
When the measurement
point is close
When measuring the thickness of the
transparent object
OUT1
HI GO LO
HI GO LO
TIM
OUT2
TIM
OUT1
HI GO LO
HI GO LO
TIM
OUT2
TIM
OUT1
HI GO LO
HI GO LO
TIM
OUT2
TIM
Note