LK-G-M-NO2-E
2 Operations and Functions during Measurement
Operations from the display panel
1
Measure the target to be used as a reference.
Assume that "
1.2345" is displayed.
2
Press the key.
The measurement value display becomes zero.
• If you enable the auto-zero function for either OUT1 or OUT2 independently, the
measurement value display mode must be set to the single display.
If you enable the auto-zero function for both OUT1 or OUT2 at the same time, the
measurement value display mode must be set to the dual display.
The measurement value at the moment when the key is released is set as
"
0.0000".
Pressing the key for three seconds cancels auto-zero.
Input from the ZERO terminal
Auto-zero is activated for OUT1 when ZERO1 (Terminal 10) on the 12-pin terminal block
receives an input and for OUT2 when ZERO2 (terminal 8) on the expansion connector
receives an input. Ensure that the turn-on time does not exceed two seconds.
NPN type : OFF when opened/ON when short-circuited with COM (No.12)
PNP type : OFF when opened/ON when the voltage is applied
The measurement value at the turning-on moment is set as "
0.0000".
If the ZERO1 terminal or the ZERO2 terminal is turned on for two seconds or more, auto-
zero is canceled.
RS-232C interface
You can send the command from the external devices by using the RS-232C interface to set or
cancel the auto-zero function.
Refer to "Switching the Measurement Value Displays" (page 2-2) for switching the
measurement value display.
Refer to "Chapter 5 RS-232C" (page 5-1).
OUT1
HI GO LO
HI GO LO
TIM
OUT2
TIM
OUT1
HI GO LO
HI GO LO
TIM
OUT2
TIM