4-45
Chapter 4 Function Settings
4
The following table shows the relationship of scaling values along with their output
ranges and resolutions.
Scaling value [μm/V] Output range Resolution ()
1 ±10μm 0.01μm
2 ±20μm 0.01μm
5 ±50μm 0.02μm
10 ±100μm 0.04μm
20 ±200μm 0.08μm
50 ±500μm 0.2μm
100 ±1mm 0.4μm
200 ±2mm 0.8μm
500 ±5mm 2μm
1000 ±10mm 4μm
2000 ±20mm 8μm
5000 ±50mm 20μm
10000 ±100mm 40μm
If the minimum display unit selected is larger than the resolution selected, the mini-
mum display unit will be used as applicable resolution.
Setting Method
1 Press the
ENT
key when the output
setting is displayed.
2 Press the
FUNC
key four times.
Measured value
0V
+10.5V
+10V
Voltage
–10V
030
–10.5V
(Positive value
over range)
(Negative value
below range)
+10V :30mm
–10V :
0mm
HI
GO
LO
3 Press the or key and set the scaling value.
4 Press the
ENT
key.
The input item is entered and the output setting display appears again.
Tolerance Mode with Threshold
Set the measurement values to be output at
10 V and –10 V respectively so that the
scaling value will be decided. The voltage
can be changed while the result waiting
status is “---.-----.” (See Page 4-47)