4-53
Chapter 4 Function Settings
4
Possible Setting Conditions
Reference Value in Tolerance Mode
Reference value + Lower limit hold value < Reference value + Upper limit hold value
Threshold Value in Tolerance Mode
Lower limit hold value < Upper limit hold value
Note: If the input conditions are not satisfied, “Error-6” will appear on the main
display. Reset the error status by pressing the
ENT
key or
ESC
key and make
the settings again.
HI
GO
LO
The lower limit
hold balue
display
HI
GO
LO
HI
GO
LO
PROG No.
PRINT
AREA DISP
ESCG
C
ENT
HOLD
ZERO
FOCUS
A
POSITION
HI
GO
LO
Press these keys to set the value.
Setting Method
Take the following steps to set the upper
limit hold value.
Press the
LO
key in steps 1 and 4 to set
the lower limit hold value.
1 Continue pressing the
HI
key for 2
seconds.
2 Press the
GO
key to change the
display to “on.”
The hold value elimination function turns
ON and the value flashes.
3 Press the or key to make set-
tings.
4 Press the
HI
key.
The input item is entered and the meas-
uring display appears again.