4-7
Chapter 4 Function Settings
4
Area Settings
Area settings are used to determine the measurements taken from a variety of objects
placed in the beam. The LS-7000 Series can use up to two measuring heads to set two
measurements (AREA1 and AREA2) for simultaneous measurements.
Area Settings
Area settings include the following items.
Area: Select the measuring head number together with
the DIA, T-EDGE, B-EDGE, or SEG mode.
Focus monitor: A function to check whether the target is within the
measuring range. Set the best measuring range
(i.e., the distance between the target and transmit-
ter) with this function.
Area check: Set the total number of edges to be measured. If
the number of edges detected does not coincide
with the set value, the measured value will be
invalidated.
Difference check: If there is a radical change between measured
values, the change will be detected and output.
Set the value for the change to be detected.
Function output selection: Set the items to be output from the output terminal
on the rear panel.
Edge detection threshold change: Set the threshold to be detected when measuring
highly transparent objects.
Area Setting Procedure
Make settings while monitoring the sub display.
The following procedure is for area settings for measuring head 1.
1 Press the
PROG
key.
PRINT
ESCPROG
FUNC
ENT
2 Check that the area to be set will
appear. Press the
ENT
key.
To set AREA2, press the
FUNC
key. Then
press the
ENT
key.
3 The area setting screen will appear.
Press the
FUNC
key to select another set
area item.
HI
GO
LO
HI
GO
LO