Chapter 6 Normal Mode
64
6
GG
GG
G Program Selection Method Setup
The program selection method can be set to the front panel or the rear terminals.
Setting Procedure
1. Select the OPTIONS screen. Press the [OPTIONS] button.
2. Select the UTILITY screen. Press the [UTILITY] button.
3. Set the program selection method.
Press the [PROGRAM SELECT] button to select [PANEL] or [EXT].
Every time you press the button, the display toggles between [PANEL] and
[EXT].
Note 1: When the program selection method is set to [PANEL], program selection
can be performed using the front panel and the RS-232C input.
Note 2: When the program selection method is set to [EXT], program selection can
be performed using the terminal input only.
Note 3: When the panel lock function is used, program selection is performed as
shown below.
For information on the panel lock function, see p. 112.
OPTIONS
UTILITY
LASER ON OPTIONS
NOR
MAL
P. 1
DISPLAY
DATA
LIMIT
SETUP
PROGRAMCALIB OPTIONS
AREA
OUTPUT
INSPECT
UTILITY
Select Option to setup.
MODE
CHANGE
Normal
-> Pitch
LASER ON UTILITY
NOR
MAL
P. 1
AREA CHECK
INITIALIZE
ON
PANEL
mm
ONE-SCAN MODE
PROGRAM SELECT
DISPLAY UNIT
ESC
Program selection
Panel Terminal
Front panel Enabled Disabled
Terminal input Disabled Enabled
RS-232C input Enabled Disabled
Program selection
Panel Terminal
Front panel Disabled Disabled
Terminal input Disabled Enabled
RS-232C input Enabled Disabled