Chapter 9 Expansion I/O Board
101
9
9.1.2 I/O Terminal Operations
11
11
1 7-level comparator output (terminals 1 to 14)
The LS-5000 series performs comparator output (
See p. 35
) for the output
number specified during the I/O slot setup (
See p. 59.
) procedure.
Note: When Pitch mode is selected, the comparator output is as follows:
HI3: Turns ON when the target’s actual number of pitches is different from the
preset number of pitches.
HI2: Unused
HI1: Turns ON when the measured value exceeds the preset upper/lower limits.
GO: Turns ON when the measured value is within the tolerance range.
LO1:Unused
LO2:Unused
LO3:Unused
22
22
2 Self-timing output (terminals 15 and 16)
In self-timing mode, the output specified during the I/O slot setup procedure turns
ON when the self-timing signal turns ON. The self-timing output is a one-shot
signal, as shown below:
See p. 59 for I/O slot setup.
33
33
3 Timing input (terminals 19 and 20)
When this terminal and the GND terminal are short-circuited, the LS-5000 series
performs the measuring mode operation for the output number specified during the
I/O slot setup procedure.
See p.46 for measuring mode and p.59 for I/O slot setup.
44
44
4 Reset input (terminals 21 and 22)
When this terminal and the GND terminal are short-circuited, the LS-5000 series
performs the measuring mode operation for the output number specified during the
I/O slot setup procedure.
See p.46 for measuring mode and p.59 for I/O slot setup.
55
55
5 Auto-zero input (terminals 23 and 22)
When this terminal and the GND terminal are short-circuited, the LS-5000 series
instantaneously resets the measured value to “0.00000” for the output number
specified during the I/O slot setup procedure.
See p.59 for I/O slot setup.
66
66
6 Analog output
The LS-5000 series performs analog output in the range of ±10 V for the output
number specified during the I/O slot setup procedure.
See p.57 for analog output and p.59 for I/O slot setup.
Response delay time
Activation of self-timing
function
50 µs
ON
OFF