Common Operations and Function Settings
5
Chapter
E LT-9001-IM
5-10
OUTPUT Settings
Normal [NORMAL]
Measures continuously, and displays and outputs the measured value as needed.
The ON status of each output means that the NPN open-collector output is ON.
The ON status of each input means that the input and the COM terminal is short-
circuited.
The RS-232C output is synchronized with the TIMING input when the "D-MODE" in the environment setting is
set to any option other than OFF. ( Page 8-17)
Refer to "Binary Output" (page 7-10) for details of the binary output.
• If setting the RESET input to ON when the TIMING input is ON, the judgment standby status is displayed ("- - -
-" is indicated) until the TIMING input becomes OFF.
• The TIMING input and RESET input can be controlled by using the RS-232C. ( Page 8-6)
The TIMING input and the RESET input can be controlled by using the [HOLD] button and the [ESCAPE]
button respectively. ( Page 5-3)
Reference
Upper limit of tolerance
Lower limit of tolerance
Measured value