5:Centerof circle
byte Setting Data
0
Arc calculation areacount:
0: Area2 disabled, 1: Area2 enabled
1 Reserved (fixed as 0)
2 Reserved (fixed as 0)
3 Reserved (fixed as 0)
4
Arc calculation area Left:
any value in measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm =500000)
5
6
7
8
Arc calculation area Right:
any value in measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm =500000)
9
10
11
12
Area2 Left: any value in measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm =500000)
13
14
15
16
Area2 Right: any value in measurement range
(0.01μm unit Sined 32-bit integer example: 5mm
=500000)
17
18
19
7:Edge (when profile compression (time axis) is off)
byte Setting Data
0
Edge measuring area Left: any value in
measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
1
2
3
4
Edge measuring area Right: any value in
measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
5
6
7
8 Edge direction: 0:Rising, 1:Falling
9 Detect direction: 0:+direction, 1:–direction
10 Detect No: 1~10
11 Reserved (fixed as 0)
12
Edge level: Any value in measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
13
14
15
7:Edge (when profile compression (time axis) is on)
byte Setting Data
0
Edge measuring area Left: any value in
measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
1
2
3
4
Edge measuring area Right: any value in
measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
5
6
7
8 Detection target: 0:Upper profile, 1:Lower profile
9 Edge direction: 0:Rising, 1:Falling
10 Detect direction: 0:+direction, 1:–direction
11 Detect No: 1~10
12
Edge level: any value in measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
13
14
15
15:P-P (X) (only for when profile compression (time axis) is
on)
byte Setting Data
0
Edge measuring area Left: any value in
measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
1
2
3
4
Edge measuring area Right: any value in
measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
5
6
7
8 Edge direction: 0:Rising, 1:Falling
9 Detect direction: 0:+direction, 1:–direction
10 Detect No: 1~10
11 Reserved (fixed as 0)
12
Edge level: any value in measurement range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
13
14
15