71LJ-V7000_COM_RM_E
<Smoothing>
Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,
・・・ , 1F:Program NO.15)
Category:02h, Item:0Bh
Target1:00h (headA/wide), 01h (headB), Target2~4:00h
byte Setting Data
0 Smoothing: 0:1, 1:2, 2:4, 3:8, 4:16, 5:32, 6:64
1 Reserved (fixed as 0)
2 Reserved (fixed as 0)
3 Reserved (fixed as 0)
<Averaging>
Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,
・・・ , 1F:Program NO.15)
Category:02h, Item:0Dh
Target1:00h (headA/wide), 01h (headB), Target2~4:00h
byte Setting Data
0
Averaging count: 0:1, 1:2, 2:4, 3:8, 4:16, 5:32,
6:64, 7:128, 8:256
1 Reserved (fixed as 0)
2 Reserved (fixed as 0)
3 Reserved (fixed as 0)
<Invalid data processing (time axis)>
Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,
・・・ , 1F:Program NO.15)
Category:02h, Item:0Eh
Target1:00h (headA/wide), 01h (headB), Target2~4:00h
byte Setting Data
0 Processing timses: 0~255
1 Resume times: 0~255
2 Reserved (fixed as 0)
3 Reserved (fixed as 0)
<Tilt correction>
Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,
・・・ , 1F:Program NO.15)
Category:02h, Item:0Fh
Target1:00h (headA/wide), 01h (headB), Target2~4:00h
byte Setting Data
0
ON/OFF: 0:Correction disabled, 1:Correction
enabled
1 Reserved (fixed as 0)
2 Reserved (fixed as 0)
3 Reserved (fixed as 0)
4
Linear calculation area count: 0:Area2 disabled,
1:Area2 enabled
5 Reserved (fixed as 0)
6 Reserved (fixed as 0)
7 Reserved (fixed as 0)
8
Area start position1: any value in measurement
range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
9
10
11
12
Area end position1: any value in measurement
range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
13
14
15
16
Area start position2: any value in measurement
range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
17
18
19
20
Area end position2: any value in measurement
range
(0.01μm unit Sined 32-bit integer
example: 5mm=500000)
21
22
23
24
Post-correction angle (–45.00~+45.00deg) :
–4500~+4500
25
26
Correction angle (–45.00~+45.00deg) :
–4500~+4500
27