Chapter 4 Use of Data Processing Functions
4
79
Flicker mode
The internal timing signal is input once at each flicker time and the "minimum
value" during the sampling period is displayed/output.
Peak-to-peak hold function [b-3]
Measures the "difference between the maximum and minimum values" during a
specified period (sampling period).
For the peak-to-peak hold function, the sampling period can be specified with the
combination of the trigger mode and delay time settings.
For the details of the trigger mode and delay time, refer to pages 69 and 70.
For the details of flicker delay, refer to pages 70 to 73.
Timing diagram
External trigger mode
The sampling period is between the UP edge of the timing input (hold-reset
input) and the UP edge of the next timing input.
The value measured at the instant when the timing input turns ON is displayed/
output.
Reference: To clear the measured value being held and to start the measurement
again, short-circuit the hold-reset and GND terminals.
(t)
0
ON
OFF
When the power is turned ON or the hold-reset
signal is input, the display shows " " until
the first measured value is determined.
Displayed value
Internal timing
Bottom hold value
Sampling period Sampling period Sampling period Sampling period
Flicker time
(t)
0
ON
OFF
p
b
p'
b'
(p-b)
(p'-b')
When the power is turned ON or the hold-reset
signal is input, the display shows " " until
the first timing signal is input.
Displayed value
Timing input
Sampling period Sampling period
Peak-to-peak
value