Chapter 4 Use of Data Processing Functions
4
76
Normal measurement function [b-0]
Displays/outputs the measured value continuously.
Timing diagram
* When the timing input is turned ON, the display and comparator output is held.
Peak hold function [b-1]
Measures the "maximum value" during a specified period (sampling period).
For the peak hold function, the sampling period can be specified with the
combination of the trigger mode and delay time settings.
For the details of the trigger mode and delay time, refer to pages 69 and 70.
For the details of flicker delay, refer to pages 70 to 73.
The sampling period is between the UP edge of the timing input (hold-reset
input) and the UP edge of the next timing input.
The value measured at the instant when the timing input turns ON is displayed/
output.
Reference: To clear the measured value being held and to start the measurement
again, short-circuit the hold-reset and GND terminals.
(t)
0
ON
OFF
ON
OFF
ON
OFF
LOW OUT
GO OUT
HIGH OUT
ON
OFF
ON
OFF
Timing input
HIGH limit value
LOW limit value
Displayed value
(t)
0
ON
OFF
When the power is turned ON or the hold-reset signal is input,
the display shows " " until the first timing signal is input.
Displayed value
Timing input
Peak hold
value
Sampling period Sampling period
Sampling period